Validation Capacity
Founded in 2007, Cobetter Filtration Validation Center has been approved by the CNAS (China National Accreditation Service for Conformity Assessment) certification. CNAS conducts laboratory tests according to the ISO 17025 system. The review, reports issued by the accredited laboratory are accepted in 58 organizations around the world.
In order to let you have a better understanding of our products of IV filter, we have done Protein binding,Retention of Pyrogenicity tests, the test results are as follows:
Protein Binding Test
Test Results
Specification | Test object | BSA (μg/cm²) | Lysozyme (μg/cm²) |
0.2μm PES 0.2+ | P company ELE96LYL | 40.86 | 6.36 |
G company 0.2+ | 36.66 | 2.42 | |
Cobetter 0.2+ | 38.10 | 2.30 | |
1.2μm PES | P company 1.2 | 0.11 | 4.77 |
G company 1.2 | 0.29 | 8.12 | |
Cobetter 1.2 | 0.15 | 1.23 |
Conclusion: Cobetter's 0.2 +, 1.2 PES membranes show similar or lower protein adsorption rates than competitors.
Retention of Pyrogenicity Test
Test Results
Solution EU/ml | Time (h) | P Company | G Company | Cobetter PES |
Challenge Solution | 0 | 1 | 1 | 1 |
8 | 1 | 1 | 1 | |
24 | 1 | 1 | 1 | |
48 | 1 | 1 | 1 | |
72 | 1 | 1 | 1 | |
96 | 1 | 1 | 1 | |
Filtrate | 0 | < 0.03 | < 0.03 | 1 |
8 | < 0.03 | < 0.03 | < 0.03 | |
24 | < 0.03 | < 0.03 | < 0.03 | |
48 | < 0.03 | < 0.03 | < 0.03 | |
72 | < 0.03 | < 0.03 | < 0.03 | |
96 | < 0.03 | < 0.03 | < 0.03 | |
Conclusion |
| Passed | Passed | Passed |
Conclusion: Conclusion: under the same comparative test conditions, cobetter 0.2 + PES filter can effectively intercept and adsorb endotoxin in 96h infusion time, which can be claimed to be suitable for 96h products